Harvestree Measurements Technical Reference

Technical reference for port measurement behavior, accuracy, alarms, and calibration by measurement type.

← Measurement ports

Conventions

  • Telemetry: values produced at the end of one measurement pass (LoRaWAN, Modbus, or local service interface, depending on device mode).
  • Changing the measurement type on a port clears that port’s calibration and alarm settings for that port.
  • Pass duration is the duration of one acquisition; how often it runs (application period, radio, sleep) is configured separately — see user Configuration / Harvestree Operation Reference.
  • Energy: figures in this document are measurement-pass budgets only (they exclude LoRaWAN radio). Per-pass use varies a lot with measurement type and settings (delay, oversampling, etc.); each section gives typical pass duration and charge examples. Radio uplink consumption and cycle scheduling are covered in Harvestree Operation Reference. For overall energy and autonomy planning, contact MOIZ or your supplier for application support.
  • Analog conditioning: most analog ports include hardware low-pass filtering on the conditioner, designed to attenuate high-frequency interference before digitization (implementation and cutoff depend on the measurement type). Further sample processing reduces what still passes the filter; see Rejection under How the reading is produced.
  • How readings are built: after a port is selected, the electronics settle, then optionally wait (delay you configure). Each section under How the reading is produced uses the same bullet labels (Settling, Rejection, Spikes or How the value is found, Still pay attention to, Calibration / Output / Behavior where applicable). Wording is harmonized within each acquisition family (DC analog, AC, digital probe, vibration, contact).

Contents

Disabled

Label Notes
Not used No measurement on the port (PORT_NONE).

Temperature

Subgroup Label Section
RTD PT1000 (RTD) § PT1000
Thermocouple Type K … Type R § Thermocouple types (K, J, T, N, S, E, B, R)
Digital Digital temperature probe (ref A) § Digital temperature probe (ref A)
Infrared Infrared (ref A) § Infrared (ref A)

Environmental

Label Section
Temperature and humidity § Temperature and humidity measurement

Voltage

Subgroup Label Section
Single-ended DC voltage § DC voltage
Single-ended AC voltage (RMS) § AC voltage, RMS and frequency
Differential DC differential voltage § DC differential voltage
Differential AC differential voltage (RMS) § AC differential voltage, RMS and frequency
Differential with external amplifier — to configure DC differential voltage with external amplifier § DC differential voltage with external amplifier
Differential with external amplifier — to configure AC differential voltage with external amplifier (RMS) § AC differential voltage with external amplifier (RMS)

Magnetic field — to configure

Label Section
DC magnetic sensor § DC magnetic sensor
AC magnetic sensor (RMS) § AC magnetic sensor (RMS)

Current — to configure

The picker exposes dedicated codes under Current; they use the same acquisition families as Magnetic field.

Subgroup Label Section
With voltage difference DC current with voltage difference § DC current with voltage difference
With voltage difference AC current with voltage difference (RMS) § AC current with voltage difference (RMS)
With magnetic field DC current with magnetic sensor § DC current with magnetic sensor
With magnetic field AC current with magnetic sensor (RMS) § AC current with magnetic sensor (RMS)

Industrial analog — to configure

Label Section
4–20 mA loop (120 Ω shunt; not +/−; loop does not power the sensor) § 4–20 mA loop
Potentiometer § Potentiometer

Vibration — to configure

Label Section
Analog 1-axis vibration (levels + FFT) § Analog 1-axis vibration

Dry contacts

Label Section
Dry contact § Dry contact (synchronous)

Under Ports, only synchronous dry contact is offered. Asynchronous edge/contact is configured with the per-port Async alarm trigger checkbox when Application mode is ALARM (not in the port measurement tree) — § Asynchronous dry-contact trigger.


Temperature

PT1000

Compatible probes

  • MOIZ-qualified Pt1000 assemblies and cabling: Catalog (RTD family). Third-party probes require written MOIZ approval.

Physical quantity

  • Temperature (°C) from a Pt1000 element via on-board excitation and conditioning, converted with the product’s Pt resistance–temperature characteristic.

Outputs and quantization

Output Type Scale
Temperature int16 0.1 °C per LSB (°C = value / 10.0)

Internal computation uses higher precision before rounding. Alarms are evaluated on the corrected temperature (after calibration), not on the rounded integer on the link.

  • 0.1 °C (telemetry packing).
  • Digitization: 12-bit ADC.

Global accuracy

  • ± 0.5 °C

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off high-frequency interference; 20–100 ms integration (oversampling) reduces supply harmonic interference.
  • Spikes — The most extreme 10% of samples in the window are not used in the average.
  • Still pay attention to — Probe contact, wiring and contact quality, slow drift during the measurement window, and interference that lasts longer than the averaging time.
  • CalibrationOffset or lookup table (manual curve) on temperature.

Typical pass duration

  • Order (20 ms + delay) + (20 ms × oversampling factor) when only the nominal timings apply (e.g. ~40 ms with minimum delay and shortest window, up to ~1.1 s with maximum delay and longest window).

Energy

  • Per pass (calculated budget): approximately 0.28 mC to 7.7 mC over the typical timing range above (equivalent to about 0.00008 to 0.00214 mAh per pass).

Alarms

  • Quantity: corrected temperature (°C).

Calibration (operator)

  • Acquisition timing: delay, oversampling (capture window).
  • Primary quantity calibration: offset or lookup table (manual curve) on temperature.

Thermocouple types (K, J, T, N, S, E, B, R)

Compatible probes

  • MOIZ-qualified thermocouple interfaceCatalog.

Physical quantity

  • Cold junction temperature and hot junction temperature (°C) for the selected thermocouple type.

Outputs and quantization

Output Type Scale
Cold junction Tcj int16 0.1 °C
Hot junction Thj int16 0.1 °C

Invalid: both 32767 if the measurement cannot be completed.

Alarms are evaluated on corrected Tcj / Thj after successful reads.

  • 0.1 °C on each output.
  • Digitization: 18 bits via dedicated circuit.

Global accuracy

  • **Tcj ± 2 °C, **Thj ± 5 °C

How the reading is produced

  • Settling — ~20 ms after the port is selected, then ~500 ms while the digital probe stabilizes; optional extra wait up to ~1 s (configurable).
  • Rejection — Brief read or communication errors: configurable retries.
  • Still pay attention to — Probe wiring, connector moisture, and ambient temperature at the cold junction (used for compensation).
  • CalibrationOffset or lookup table (manual curve) on cold and hot junction separately.

Typical pass duration

  • Dominated by the ~500 ms stabilization plus configured delay — typically ≥ ~0.5 s before telemetry.

Energy

  • Per pass (calculated budget): approximately 4.5 mC to 13.5 mC nominal (equivalent to about 0.00125 to 0.00375 mAh per successful read).
  • Worst case: up to about the upper nominal figure when all configured retries are used before success or failure.

Alarms

  • Quantities: corrected Tcj and Thj (°C).

Calibration

  • Acquisition timing: delay (extra stabilization) and retries (bus robustness).
  • Primary quantity calibration (Tcj / Thj): offset or lookup table (manual curve).

Digital temperature probe (ref A)

Compatible probes

  • MOIZ-qualified numeric temperature probe ref ACatalog.

Physical quantity

  • Temperature (°C) from the digital sensor bus.

Outputs and quantization

Output Type Scale
Temperature int16 0.1 °C

Invalid: 32767; alarms are not asserted on a failed read.

Alarms are evaluated on corrected temperature after a successful read.

  • 0.1 °C.
  • Digitization: 16 bits via dedicated circuit.

Global accuracy

  • ± 0.5 °C

How the reading is produced

  • Settling — ~50 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • Rejection — Brief read or communication errors: configurable retries.
  • Still pay attention to — Cable length, supply quality at the probe, and sensor self-heating on fast repeats.
  • CalibrationOffset or lookup table (manual curve) on temperature.

Typical pass duration

  • ~55 ms + delay in the nominal case.

Energy

  • Per pass (calculated budget): approximately 0.385 mC to 7.4 mC nominal (equivalent to about 0.00011 to 0.00206 mAh per successful read).
  • Worst case: up to about the upper nominal figure when all configured retries are used before success or failure.

Alarms

  • Quantity: corrected temperature (°C) after a successful sensor read.

Calibration

  • Acquisition timing: delay (extra stabilization) and retries (bus robustness).
  • Primary quantity calibration (temperature): offset or lookup table (manual curve).

Infrared (ref A)

Compatible probes

  • MOIZ-qualified infrared temperature interfaceCatalog.

Physical quantity

  • Object temperature (°C).

Outputs and quantization

Output Type Scale
Object temperature int16 0.1 °C

Invalid: 32767; alarms are not asserted on a failed read.

Alarms are evaluated on corrected object temperature after a successful read.

  • 0.1 °C on the link. In practice, measurement error mainly depends on installation conditions (field of view, target distance, and emissivity setting).
  • Digitization: 17-bit ADC via dedicated circuit.

Global accuracy

  • ± 1 °C

How the reading is produced

  • Settling — Sensor power is applied, then ~1 s warm-up before the first valid reading; optional extra wait up to ~1 s (configurable).
  • Rejection — Brief read or communication errors: configurable retries.
  • Still pay attention to — Field of view, distance, surface emissivity, and reflections from hot or shiny surroundings.
  • CalibrationEmissivity when configured; offset or lookup table (manual curve) on object temperature.

Typical pass duration

  • ≥ ~1 s + delay in normal conditions.

Energy

  • Per pass (calculated budget): approximately 9 mC to 18 mC nominal (equivalent to about 0.0025 to 0.005 mAh per successful read).
  • Worst case: up to about the upper nominal figure when all configured retries are used before success or failure.

Alarms

  • Quantity: corrected object temperature (°C) after a successful sensor read.

Calibration

  • Acquisition timing: delay (extra stabilization) and retries (sensor read robustness).
  • Primary quantity calibration (object temperature): emissivity setting; offset or lookup table (manual curve).

Environmental

Temperature and humidity measurement

Compatible modules

  • MOIZ-qualified temperature/humidity module ref ACatalog.

Physical quantity

  • Temperature (°C) and relative humidity (%RH). Use this measurement for standard environmental monitoring. A frost index may also be reported (derived from T and RH) as an optional icing-risk indicator — not required when you only need temperature and humidity.

Outputs and quantization

Output Type Scale
Temperature int16 0.1 °C
Humidity int16 0.1 %RH
Frost index (optional) uint8 Derived frost index (from T and RH)

Invalid: T and H may be 32767; frost is set to 0 on failed read.

Alarms are evaluated on corrected T and RH after a successful read; the derived frost index can also be used when it is produced.

  • 0.1 °C, 0.1 %RH; frost index is coarser.
  • Digitization: 16 bits via dedicated circuit.

Global accuracy

  • T ± 1 °C; RH ± 1 %RH

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • Rejection — Brief read or communication errors: configurable retries.
  • Condensation — At high humidity and low temperature, an internal heater cycle may run and add ~1 s before a new sample (worst case > 1 s per pass).
  • Still pay attention to — Sensor exposure (rain, direct sun), mounting airflow, and self-heating from the heater cycle on repeated passes.
  • CalibrationOffset or lookup table (manual curve) on temperature and humidity; frost indicator uses corrected values when enabled.

Typical pass duration

  • Often tens to hundreds of ms + delay; > 1 s possible when anti-condensation logic runs.

Energy

  • Per pass (calculated budget): approximately 0.35 mC to 8.4 mC nominal (equivalent to about 0.00010 to 0.00233 mAh per successful read).
  • Worst case: up to about the upper nominal figure when all configured retries are used before success or failure; anti-condensation heater cycles can add further energy (see below).
  • With anti-condensation heater active, budget is approximately 10 mC to 22 mC (about 0.00278 to 0.00611 mAh) depending on added heater/re-sampling timing.

Alarms

  • Quantities: T and RH (corrected on the link); optional frost index when derived from a successful read.

Calibration

  • Acquisition timing: delay (extra stabilization) and retries (bus robustness).
  • Primary quantity calibration (temperature): offset or lookup table (manual curve).
  • Primary quantity calibration (humidity): offset or lookup table (manual curve).

Voltage

DC voltage

DC voltage uses the DC acquisition path and transmits a scaled integer value.

Compatible probes

  • MOIZ-qualified voltage assemblies and cabling for single-ended DC within the qualified input range: Catalog. Third-party assemblies or cabling require written MOIZ approval.

Physical quantity

  • Voltage at the conditioned input (mV before user correction).

Outputs and quantization

Output Type Scale
Voltage int16 0.1 mV

Alarms are evaluated on corrected mV (after calibration), not on raw ADC counts.

  • 0.1 mV (int16).
  • Digitization: 12-bit ADC.

Global accuracy

  • ± 1.5 mV

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off high-frequency interference; 20–100 ms integration (oversampling) reduces supply harmonic interference.
  • Spikes — The most extreme 10% of samples in the window are not used in the average.
  • Still pay attention to — Wiring and contact quality, slow drift during the measurement window, and interference that lasts longer than the averaging time.
  • CalibrationOffset or lookup table (manual curve) on the voltage channel.

Typical pass duration

  • (20 ms + delay) + (20 ms × oversampling factor).

Energy

  • Per pass (calculated budget): approximately 0.28 mC to 7.7 mC over the typical timing range above (equivalent to about 0.00008 to 0.00214 mAh per pass).

Alarms

  • Quantity: corrected voltage (mV) at the conditioned input.

Calibration

  • Acquisition timing: delay and oversampling (capture window).
  • Primary quantity calibration (voltage): offset or lookup table (manual curve).

AC voltage, RMS and frequency

Compatible probes

  • MOIZ-qualified voltage assemblies and cabling for single-ended AC within the qualified input range: Catalog. Third-party assemblies or cabling require written MOIZ approval.

Physical quantity

  • Fundamental RMS (mV) and fundamental frequency (Hz) estimated in the product’s low-frequency band.

Outputs and quantization

Output Type Scale
RMS int16 0.1 mV RMS
Frequency int16 0.1 Hz

Alarms are evaluated on corrected RMS and corrected frequency (after calibration).

  • 0.1 mV RMS and 0.1 Hz (int16).
  • Digitization: 12-bit ADC.

Global accuracy

  • RMS ± 1.5 mV; frequency ± 1 Hz

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off frequencies outside the AC measurement band; 100 ms or 200 ms acquisition estimates the dominant tone in 20–100 Hz.
  • How the value is found — DC offset is removed, then a single-frequency RMS is computed at the dominant frequency (Goertzel-based) with front-end correction.
  • Still pay attention to — Distorted waveforms, multiple strong frequencies in band, and very low signal level.
  • Calibration — RMS and frequency calibrated separately.

Typical pass duration

  • (20 ms + delay) + (100 ms or 200 ms) acquisition.

Energy

  • Per pass (calculated budget): approximately 0.84 mC to 8.54 mC over the typical timing range above (equivalent to about 0.00023 to 0.00237 mAh per pass).

Alarms

  • Quantities: corrected fundamental RMS (mV) and corrected fundamental frequency (Hz) from the narrow-band estimate in How the reading is produced.

Calibration

  • Acquisition timing: analysis window (100 ms or 200 ms) and delay.
  • RMS quantity calibration: offset or lookup table (manual curve).
  • Frequency quantity calibration: offset or lookup table (manual curve).

DC differential voltage

Compatible probes

  • MOIZ-qualified differential voltage assemblies and cabling for differential DC within the qualified common-mode and differential range (built-in path): Catalog. Third-party assemblies or cabling require written MOIZ approval.

Physical quantity

  • Differential voltage (mV).

Outputs and quantization

Output Type Scale
Differential voltage int16 0.1 mV

Alarms are evaluated on corrected differential mV (after calibration).

  • 0.1 mV (int16) for differential voltage.
  • Digitization: 12-bit ADC.

Global accuracy

  • ± 0.5 mV

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off high-frequency interference; 20–100 ms integration (oversampling) reduces supply harmonic interference.
  • Spikes — The most extreme 10% of samples in the window are not used in the average.
  • Still pay attention to — Wiring and contact quality, slow drift during the measurement window, and interference that lasts longer than the averaging time.
  • Output — Differential chain with fixed onboard gain, reported as differential mV.
  • CalibrationOffset or lookup table (manual curve) on differential voltage.

Typical pass duration

  • (20 ms + delay) + (20 ms × oversampling factor).

Energy

  • Per pass (calculated budget): approximately 0.28 mC to 7.7 mC over the typical timing range above (equivalent to about 0.00008 to 0.00214 mAh per pass).

Alarms

  • Quantity: corrected differential voltage (mV) after onboard gain and correction.

Calibration

  • Acquisition timing: delay and oversampling (capture window).
  • Primary quantity calibration (differential voltage): offset or lookup table (manual curve).

AC differential voltage, RMS and frequency

Compatible probes

  • MOIZ-qualified differential voltage assemblies and cabling for differential AC within the qualified common-mode and differential range (built-in path): Catalog. Third-party assemblies or cabling require written MOIZ approval.

Physical quantity

  • Fundamental RMS (mV) and frequency (Hz) for the differential pair.

Outputs and quantization

Output Type Scale
RMS int16 0.1 mV RMS
Frequency int16 0.1 Hz

Alarms are evaluated on corrected RMS and corrected frequency (after calibration).

  • 0.1 mV RMS and 0.1 Hz (int16).
  • Digitization: 12-bit ADC.

Global accuracy

  • RMS ± 0.5 mV; frequency ± 1 Hz

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the differential conditioner rolls off frequencies outside the AC measurement band; 100 ms or 200 ms acquisition estimates the dominant tone in 20–100 Hz.
  • How the value is found — DC offset is removed, then a single-frequency RMS is computed at the dominant frequency (Goertzel-based) with front-end correction.
  • Still pay attention to — Wiring and contact quality, distorted waveforms, multiple strong frequencies in band, and very low signal level.
  • Calibration — RMS and frequency calibrated separately.

Typical pass duration

  • (20 ms + delay) + 100 ms or 200 ms.

Energy

  • Per pass (calculated budget): approximately 0.84 mC to 8.54 mC over the typical timing range above (equivalent to about 0.00023 to 0.00237 mAh per pass).

Alarms

  • Quantities: corrected fundamental RMS (mV) and frequency (Hz) on the differential AC path (same estimator as single-ended AC, different RC corner — Resolution on the link).

Calibration

  • Acquisition timing: analysis window (100 ms or 200 ms) and delay.
  • RMS quantity calibration: offset or lookup table (manual curve).
  • Frequency quantity calibration: offset or lookup table (manual curve).

DC differential voltage with external amplifier

Compatible probes

  • MOIZ-qualified external amplifiers and cabling for differential DC, with packaged gains: Catalog. Third-party amplifiers or cabling require written MOIZ approval.

Physical quantity

  • Differential voltage in configured engineering units (default: mV).

Outputs and quantization

Output Type Scale
Primary quantity float32 Engineering unit after calibration (default: mV)

Alarms are evaluated on corrected primary quantity (after calibration).

  • float32 primary quantity.
  • Digitization: 12-bit ADC.

Global accuracy

  • ± 0.1 %FS (measurement chain) + external amplifier accuracy.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off high-frequency interference; 20–100 ms integration (oversampling) reduces supply harmonic interference.
  • Spikes — The most extreme 10% of samples in the window are not used in the average.
  • Still pay attention to — External amplifier gain, wiring and contact quality, slow drift during the measurement window, and interference that lasts longer than the averaging time.
  • Output — Value scaled to configured engineering units (default mV) using amplifier gain and calibration.
  • CalibrationExternal amplifier gain preset (packaged lookup table).

Typical pass duration

  • (20 ms + delay) + (20 ms × oversampling factor).

Energy

  • Per pass (calculated budget): approximately 0.32 mC to 8.96 mC over the typical timing range above (equivalent to about 0.00009 to 0.00249 mAh per pass).

Alarms

  • Quantity: corrected primary quantity in configured engineering units (default: mV).

Calibration

  • Delay and oversampling (capture window control).
  • Primary quantity calibration: selection of external amplifier gain

AC differential voltage with external amplifier (RMS)

Compatible probes

  • MOIZ-qualified external amplifiers and cabling for differential AC, with packaged gains: Catalog. Third-party amplifiers or cabling require written MOIZ approval.

Physical quantity

  • Fundamental RMS in configured engineering RMS units (default: mV RMS) and corrected frequency (Hz).

Outputs and quantization

Output Type Scale
RMS float32 Engineering RMS unit after calibration (default: mV RMS)
Frequency float32 Hz

Alarms are evaluated on corrected RMS and corrected frequency (after calibration).

  • float32 RMS and float32 frequency.
  • Analysis window: 100 ms or 200 ms.
  • Digitization: 12-bit ADC.

Global accuracy

  • RMS ± 0.1 %FS (measurement chain) + external amplifier accuracy; frequency ± 1 Hz.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner (external amplifier in path) rolls off frequencies outside the AC measurement band; 100 ms or 200 ms acquisition estimates the dominant tone in 20–100 Hz.
  • How the value is found — DC offset is removed, then a single-frequency RMS is computed at the dominant frequency (Goertzel-based) with front-end and external amplifier gain correction.
  • Still pay attention to — External amplifier gain, distorted waveforms, multiple strong frequencies in band, and very low signal level.
  • CalibrationExternal amplifier gain preset (packaged lookup table) on RMS; frequency not calibrated.

Typical pass duration

  • (20 ms + delay) + 100 ms or 200 ms.

Energy

  • Per pass (calculated budget): approximately 0.96 mC to 9.76 mC over the typical timing range above (equivalent to about 0.00027 to 0.00271 mAh per pass).

Alarms

  • Quantities: corrected RMS in configured engineering units and corrected frequency.

Calibration

  • Analysis window (100 ms or 200 ms) and delay before sampling.
  • RMS quantity calibration: selection of external amplifier gain
  • Frequency quantity calibration: none

Magnetic field

DC magnetic sensor

Compatible use

  • MOIZ-qualified magnetic field sensors (Hall, TMR, and other catalog-listed types) — Catalog.

Physical quantity

  • Magnetic field expressed in configured engineering units after calibration.

Outputs and quantization

Output Type Scale
Primary quantity float32 Engineering unit after calibration
  • float32 primary quantity.

Global accuracy

  • ± 0.1 %FS (measurement chain) + sensor %FS.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off high-frequency interference; 20–100 ms integration (oversampling) reduces supply harmonic interference.
  • Spikes — The most extreme 10% of samples in the window are not used in the average.
  • Still pay attention to — Sensor gap, target speed, magnetic background fields, slow drift during the measurement window, and interference that lasts longer than the averaging time.
  • CalibrationMagnetic sensor type preset (packaged lookup table).

Typical pass duration

  • (20 ms + delay) + (20 ms × oversampling factor).

Energy

  • Per pass (calculated budget): approximately 0.6 mC to 16.8 mC over the typical timing range above (equivalent to about 0.00017 to 0.00467 mAh per pass).

Alarms

  • Quantity: corrected magnetic-field quantity in configured units.

Calibration

  • Delay and oversampling (capture window control).
  • Primary quantity calibration: selection of magnetic sensor type preset (packaged lookup table)

AC magnetic sensor (RMS)

Compatible use

  • MOIZ-qualified magnetic field sensors (Hall, TMR, and other catalog-listed types) — Catalog.

Physical quantity

  • Magnetic-field RMS and frequency.

Outputs and quantization

Output Type Scale
RMS float32 Engineering RMS unit after calibration
Frequency float32 Hz
  • float32 RMS and float32 frequency.
  • Analysis window: 100 ms or 200 ms.

Global accuracy

  • ± 0.1 %FS (measurement chain) + sensor %FS.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the magnetic front-end rolls off frequencies outside the AC measurement band; 100 ms or 200 ms acquisition estimates the dominant tone in 20–100 Hz.
  • How the value is found — DC offset is removed, then a single-frequency RMS is computed at the dominant frequency (Goertzel-based) with front-end correction.
  • Still pay attention to — Sensor alignment, speed variation, multiple strong frequencies in band, and very low signal level.
  • Calibration — RMS and frequency calibrated separately.

Typical pass duration

  • (20 ms + delay) + 100 ms or 200 ms.

Energy

  • Per pass (calculated budget): approximately 1.8 mC to 18.3 mC over the typical timing range above (equivalent to about 0.00050 to 0.00508 mAh per pass).

Alarms

  • Quantities: corrected RMS in configured engineering units and corrected frequency.

Calibration

  • Analysis window (100 ms or 200 ms) and delay before sampling.
  • RMS quantity calibration: selection of magnetic sensor type preset (packaged lookup table)
  • Frequency quantity calibration: none

Current

DC current with voltage difference

Compatible probes

  • MOIZ-qualified differential voltage assemblies and cabling for differential DC within the qualified common-mode and differential range (built-in path): Catalog. Third-party assemblies or cabling require written MOIZ approval.

Physical quantity

  • Current expressed in configured engineering units after calibration.

Outputs and quantization

Output Type Scale
Primary quantity float32 Engineering unit after calibration
  • float32 primary quantity.

Global accuracy

  • ± 0.1 %FS (measurement chain) + external contribution.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off high-frequency interference; 20–100 ms integration (oversampling) reduces supply harmonic interference.
  • Spikes — The most extreme 10% of samples in the window are not used in the average.
  • Still pay attention to — Loop wiring, burden voltage, slow drift during the measurement window, and interference that lasts longer than the averaging time.
  • Output — Current derived from measured differential voltage.
  • CalibrationLinear mapping (guided parameters) or lookup table (manual curve) on the current channel.

Typical pass duration

  • (20 ms + delay) + (20 ms × oversampling factor).

Energy

  • Per pass (calculated budget): approximately 0.6 mC to 16.8 mC over the typical timing range above (equivalent to about 0.00017 to 0.00467 mAh per pass).

Alarms

  • Quantity: corrected current quantity in configured units.

Calibration

  • Delay and oversampling (capture window control).
  • Primary quantity calibration: linear mapping (from zero-point raw, linear resistivity, and distance between points) or lookup table (manual curve)

AC current with voltage difference (RMS)

Compatible probes

  • MOIZ-qualified current-sensing assemblies and cabling for differential AC pickup on the conductor, within the qualified common-mode and differential range (built-in path): Catalog. Third-party assemblies or cabling require written MOIZ approval.

Physical quantity

  • Current RMS and frequency.

Outputs and quantization

Output Type Scale
RMS float32 Engineering RMS unit after calibration
Frequency float32 Hz
  • float32 RMS and float32 frequency.
  • Analysis window: 100 ms or 200 ms.

Global accuracy

  • RMS ± 0.1 %FS (measurement chain) + external contribution; frequency ± 1 Hz.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the differential conditioner rolls off frequencies outside the AC measurement band; 100 ms or 200 ms acquisition estimates the dominant tone in 20–100 Hz.
  • How the value is found — DC offset is removed, then a single-frequency RMS is computed at the dominant frequency (Goertzel-based) with front-end correction.
  • Still pay attention to — Loop wiring, burden voltage, distorted waveforms, multiple strong frequencies in band, and very low signal level.
  • Output — Current derived from measured differential voltage.
  • Calibration — RMS and frequency calibrated separately.

Typical pass duration

  • (20 ms + delay) + 100 ms or 200 ms.

Energy

  • Per pass (calculated budget): approximately 1.8 mC to 18.3 mC over the typical timing range above (equivalent to about 0.00050 to 0.00508 mAh per pass).

Alarms

  • Quantities: corrected RMS in configured engineering units and corrected frequency.

Calibration

  • Analysis window (100 ms or 200 ms) and delay before sampling.
  • RMS quantity calibration: linear mapping (from zero-point raw, linear resistivity, and distance between points) or lookup table (manual curve)
  • Frequency quantity calibration: offset or lookup table (manual curve)

DC current with magnetic sensor

Compatible use

  • MOIZ-qualified current sensors using a magnetic field pickup (Hall, TMR, and other catalog-listed types) — Catalog.

Physical quantity

  • Current expressed in configured engineering units after calibration.

Outputs and quantization

Output Type Scale
Primary quantity float32 Engineering unit after calibration
  • float32 primary quantity.

Global accuracy

  • ± 0.1 %FS (measurement chain) + sensor %FS.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off high-frequency interference; 20–100 ms integration (oversampling) reduces supply harmonic interference.
  • Spikes — The most extreme 10% of samples in the window are not used in the average.
  • Still pay attention to — Sensor gap, target speed, magnetic background fields, slow drift during the measurement window, and interference that lasts longer than the averaging time.
  • CalibrationMagnetic sensor type preset (packaged lookup table).

Typical pass duration

  • (20 ms + delay) + (20 ms × oversampling factor).

Energy

  • Per pass (calculated budget): approximately 1.8 mC to 18.3 mC over the typical timing range above (equivalent to about 0.00050 to 0.00508 mAh per pass).

Alarms

  • Quantity: corrected current quantity in configured units.

Calibration

  • Delay and oversampling (capture window control).
  • Primary quantity calibration: linear mapping (from zero-point raw and coupling factor (A/mV)) or lookup table (manual curve)

AC current with magnetic sensor (RMS)

Compatible use

  • MOIZ-qualified current sensors using a magnetic field pickup (Hall, TMR, and other catalog-listed types) — Catalog.

Physical quantity

  • Current RMS and frequency.

Outputs and quantization

Output Type Scale
RMS float32 Engineering RMS unit after calibration
Frequency float32 Hz
  • float32 RMS and float32 frequency.
  • Analysis window: 100 ms or 200 ms.

Global accuracy

  • ± 0.1 %FS (measurement chain) + sensor %FS.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the magnetic front-end rolls off frequencies outside the AC measurement band; 100 ms or 200 ms acquisition estimates the dominant tone in 20–100 Hz.
  • How the value is found — DC offset is removed, then a single-frequency RMS is computed at the dominant frequency (Goertzel-based) with front-end correction.
  • Still pay attention to — Sensor alignment, speed variation, multiple strong frequencies in band, and very low signal level.
  • Output — Current derived from the magnetic pickup.
  • Calibration — RMS and frequency calibrated separately.

Typical pass duration

  • (20 ms + delay) + 100 ms or 200 ms.

Energy

  • Per pass (calculated budget): approximately 1.8 mC to 18.3 mC over the typical timing range above (equivalent to about 0.00050 to 0.00508 mAh per pass).

Alarms

  • Quantities: corrected RMS in configured engineering units and corrected frequency.

Calibration

  • Analysis window (100 ms or 200 ms) and delay before sampling.
  • RMS quantity calibration: linear mapping (from zero-point raw and coupling factor (A/mV)) or lookup table (manual curve)
  • Frequency quantity calibration: offset or lookup table (manual curve)

Industrial analog

4–20 mA loop

Compatible probes

  • MOIZ-qualified 4–20 mA loop cabling with embedded 120 Ω shunt for externally powered loops (device does not power the loop): Catalog. Cable use is required for compatibility. Third-party cabling or loop wiring require written MOIZ approval.
  • If the embedded 120 Ω burden is insufficient for the selected 4–20 mA sensor/transmitter, add external loop load as required by the sensor compliance specification.

Physical quantity

  • Primary quantity in configured engineering units (default: mA from the shunt current).

Outputs and quantization

Output Type Scale
Current float32 Engineering unit after calibration (default: mA)

Alarms are evaluated on corrected primary quantity (after calibration).

  • float32 primary quantity.
  • Digitization: 12-bit ADC.

Global accuracy

  • ± 0.02 mA (measurement chain) + sensor/transmitter accuracy.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off high-frequency interference; 20–100 ms integration (oversampling) reduces supply harmonic interference.
  • Spikes — The most extreme 10% of samples in the window are not used in the average.
  • Still pay attention to — Loop supply stability, wiring resistance, ground differences, slow drift during the measurement window, and interference that lasts longer than the averaging time.
  • Output — Current from shunt voltage (I = V / 120 Ω).
  • CalibrationNone (raw shunt-derived mA), linear mapping (guided parameters), or lookup table (manual curve).

Typical pass duration

  • (20 ms + delay) + (20 ms × oversampling factor).

Energy

  • Per pass (calculated budget): approximately 0.28 mC to 7.84 mC over the typical timing range above (equivalent to about 0.00008 to 0.00218 mAh per pass).

Alarms

  • Quantity: corrected primary quantity in configured engineering units (default: mA).

Calibration

  • Delay and oversampling (capture window control).
  • Primary quantity calibration: none (raw shunt-derived mA), linear mapping (two anchor points on sensor envelope), or lookup table (manual curve)

Potentiometer

Compatible probes

  • MOIZ-qualified potentiometer assemblies and cabling within the qualified input range: Catalog. Third-party assemblies or cabling require written MOIZ approval.

Physical quantity

  • Primary quantity in configured engineering units (default: % of span from the potentiometer ratio).

Outputs and quantization

Output Type Scale
Position float32 Engineering unit after calibration (default: % of span)

Alarms are evaluated on corrected primary quantity (after calibration).

  • float32 primary quantity.
  • Digitization: 12-bit ADC.

Global accuracy

  • ± 0.1 %FS (measurement chain) + sensor accuracy.

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionRC low-pass on the conditioner rolls off high-frequency interference; 20–100 ms integration (oversampling) reduces supply harmonic interference.
  • Spikes — The most extreme 10% of samples in the window are not used in the average.
  • Still pay attention to — Wiring and contact quality, slow drift during the measurement window, and interference that lasts longer than the averaging time.
  • Output — Reported as % of span.
  • CalibrationLinear mapping (guided parameters) or lookup table (manual curve) on the primary quantity.

Typical pass duration

  • (20 ms + delay) + (20 ms × oversampling factor).

Energy

  • Per pass (calculated budget): approximately 0.28 mC to 7.7 mC over the typical timing range above (equivalent to about 0.00008 to 0.00214 mAh per pass).

Alarms

  • Quantity: corrected primary quantity in configured engineering units (default: % of span).

Calibration

  • Delay and oversampling (capture window control).
  • Primary quantity calibration: linear mapping (from zero-point raw (% FS), engineering range max, and unit selection) or lookup table (manual curve)

Vibration

Analog 1-axis vibration

Compatible use

  • Single-axis vibration measurement; compatible products are listed in the Catalog.

Physical quantity

  • LF: spectral RMS as velocity (mm/s) over 10 Hz–1 kHz.
  • HF: spectral RMS as acceleration (m/s²) over 2 kHz–10 kHz.
  • Sub-band energy ratios (LF three, HF two) for relative band content.

Outputs and quantization

Field Type Meaning
RMS LF int16 LF RMS × 100 (mm/s)
RMS HF int16 HF RMS × 100 (m/s²)
LF / HF ratio bytes uint8 Energy share × 255

Ratios are telemetry only (not alarm inputs).

Alarms apply only to LF and HF RMS (corrected), not to ratio bytes.

  • LF 3.2 kHz, HF 32 kHz; 2048-point spectra after warm-up; Hann window. Telemetry scaling as in the table above.

Global accuracy

  • LF RMS ± 5 %; HF RMS ± 5 %

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • RejectionLow-pass on the vibration path (cutoff tracks sampling rate) rolls off content above the useful acquisition band.
  • How the value is found — Spectral integration reports energy in configured low- and high-frequency ranges.
  • Still pay attention to — Mounting, cable routing, temperature, and band limits vs. the fault you are monitoring.
  • Calibration — Sensitivity, offset, optional band edges and delay; optional time trace or spectrum in host software.

Typical pass duration

  • (20 ms + delay) + ~1 s spectral acquisition (LF then HF bands).

Energy

  • Per pass (calculated budget): approximately 10 mC to 20 mC over the typical timing range above (equivalent to about 0.00278 to 0.00556 mAh per pass).

Alarms

  • Quantities: corrected LF RMS and corrected HF RMS in configured engineering units.

Calibration

  • Acquisition timing: delay before spectral acquisition.
  • Primary quantity calibration (overall channel): sensitivity and offset
  • Spectral partitioning: LF / HF sub-band edges overrides

Dry contacts

Dry contact (synchronous)

Physical quantity

  • Contact state (open / closed) with defined excitation on the port.

Outputs and quantization

Output Type Meaning
State uint8 0 = open, 1 = closed (unless invert logical state is enabled in calibration)

How the reading is produced

  • Settling — ~20 ms after the port is selected; optional extra wait up to ~1 s (configurable).
  • Rejection — Hardware filtering and debounce on the contact path.
  • Output — One sampled open / closed state per pass; logic can be inverted in calibration.
  • Still pay attention to — Contact bounce, wet or high-resistance contacts, and cable capacitance on long runs.

Global accuracy

  • N/A (discrete open / closed; alarms on 0/1 logic)

Typical pass duration

  • ~20 ms + delay (no multi-sample averaging).

Energy

  • Per pass (calculated budget): approximately 0.1 mC to 5.1 mC over the typical timing range above (equivalent to about 0.00003 to 0.00142 mAh per pass).

Alarms

  • Quantity: open / closed state, exposed to the threshold engine as a numeric value (0/1 or equivalent) for High / Low / In range / Out of range rules.

Calibration

  • Acquisition timing: sense delay
  • Primary quantity calibration: invert logical state

Dry contacts — asynchronous

Asynchronous dry-contact trigger

This path is a contact trigger for ALARM mode, not a scheduled measurement type like the ports above.

Purpose

  • Wake / trigger the product on a dry-contact edge (rising or falling), then monitor the contact until it returns to normal.
  • For a sampled open/closed value on each application period, use § Dry contact (synchronous) instead.

Availability and configuration

  • Usable only when application mode is ALARM (not in normal periodic measurement scheduling).
  • Per port: enable the async alarm trigger and choose rising or falling edge (plus alarm flags as needed).
  • No unified port calibration, no logical inversion, and no per-port measurement alarm table like synchronous dry contact.

Output

  • When read for telemetry or servicing: uint8 0 = open, 1 = closed (same encoding as synchronous dry contact).

How the reading is produced

  • Rejection — Hardware filtering and debounce on the contact input.
  • Behavior — The configured edge into the alarm-active contact state starts the ALARM flow; that state is then checked on each alarm-cycle wake until the contact is back to normal (no re-trigger on further edges during the episode).
  • Still pay attention to — Noise on long cables, supply coupling, and false triggers from vibration or EMI.

Energy

  • Passive path current on the contact circuit: approximately 1 µA open, 330 µA closed.